IJ SERIES
INTERMEDIATE JUNCTION
Instruction Manual
IJ 44 A/C - pH electrode - spear tip
IJ 40 A/C - pH electrode- spear tip
IJ 64 - ORP/Redox electrode
IJ 46 - pH electrode - at membrane
IJ 14 - Reference electrode
IJ Ag - Silver Billet electrode
IJ Au - Gold ORP electrode
Including High Temperature (HT)
and Temperature Compensated (TC)Models
12 Walker St. Tennyson Qld. 4105
+ 617 3848 1660
www.ionode.com
V5.1
Przeglądanie stron 0
1 2 3 4 5 6 ... 43 44

Podsumowanie treści

Strona 1 - IJ SERIES

IJ SERIESINTERMEDIATE JUNCTIONInstruction ManualIJ 44 A/C - pH electrode - spear tipIJ 40 A/C - pH electrode- spear tipIJ 64 - ORP/Re

Strona 2 - Table of Contents

Ionode IJ Instruction Manual 10 The reference half-cell completes the circuit. The primary IJ reference is a silver/silver chloride elem

Strona 3 - 2. Parts List

Ionode IJ Instruction Manual 11 If the restriction becomes clogged and movement of ions becomes inhibited, the electrode will appear to

Strona 4 - 3. Preparation

Ionode IJ Instruction Manual 12 Figure 11 Not to Scale

Strona 5

Ionode IJ Instruction Manual 13 Offset Each pH electrode has a small offset which varies from electrode to electrode. It is due to an

Strona 6 - 4.1. Theory

Ionode IJ Instruction Manual 14 calibration. It can also be accurately measured using the mV mode of a pH meter. 1) Measure the temperature of two

Strona 7 - 4.2.1. Construction

Ionode IJ Instruction Manual 15 further details, please contact Ionode specifying make and model of pH/mV meter or titrator. Sodium Error The se

Strona 8

Ionode IJ Instruction Manual 16 Graph 1

Strona 9

Ionode IJ Instruction Manual 17 Response Time This varies according to the pH glass composition, membrane wall thickness, pH meter filteri

Strona 10

Ionode IJ Instruction Manual 18 4.3. Measurement 1) Turn on the pH meter and connect the pH electrode. Connect a temperature sensor (ATC) i

Strona 11 - 4.2.2. Output

Ionode IJ Instruction Manual 19 4.3.1. Helpful Hints 1) Use the same technique to measure samples as was used for calibration. Be consi

Strona 12 - Not to Scale

Table of Contents 1. Introduction 3 2. Parts List 3 3. Preparation 4 4. pH 6 4.1. Theory 6 4.2. Electrode Characteristics 7 4.3. Measurem

Strona 13 - increases as the

Ionode IJ Instruction Manual 20 4.4. Maintenance 1) When not in use, keep the electrode immersed in 2-3M KCl (saturated KCl is approx

Strona 14

Ionode IJ Instruction Manual 21 4.4.1. Cleaning As pH and ORP are surface based measurements, the cleanliness of the sensor and junctio

Strona 15

Ionode IJ Instruction Manual 22 4.5. Calibration Standards 4.5.1. NBS pH Buffers pH 4.01: 0.05M Potassium Hydrogen Phthalate Dissolve 10.12g of

Strona 16

Ionode IJ Instruction Manual 23 4.5.2. Secondary pH Buffers pH 7.65: Tris Buffer Dry Tris (2-amino-2-(hydroxymethyl)propane-1,3-diol) and Tris

Strona 17

Ionode IJ Instruction Manual 24 Table 2: pH Values of Standard Buffers at Various Temperatures Temperature (T) 0.05M potassium hydrogen phth

Strona 18 - 4.3. Measurement

Ionode IJ Instruction Manual 25 4.5.3. Sleeve Electrolytes The IJ reference system allows a variety of sleeve electrolytes. For high visco

Strona 19 - 4.3.1. Helpful Hints

Ionode IJ Instruction Manual 26 3M Sodium Chloride Add 17.5g of NaCl to 100 mL volumetric flask and fill to the mark with distilled water.

Strona 20 - 4.4. Maintenance

Ionode IJ Instruction Manual 27 4.6. Troubleshooting A fault may develop in either the meter or electrode. If a number of electrodes a

Strona 21 - 4.4.1. Cleaning

Ionode IJ Instruction Manual 28 Symptom Possible Causes Remedy Low Slope (<94%) Contaminated Buffers Replace buffers Membrane not Clean See 4.6

Strona 22 - 4.5. Calibration Standards

Ionode IJ Instruction Manual 29 Symptom Possible Causes Remedy Large Offset (>0.5pH) Insufficient Sleeve Electrolyte See 3. Preparation Junction

Strona 23 - 4.5.2. Secondary pH Buffers

Ionode IJ Instruction Manual 3 Figure 1 1. Introduction The IJ series of electrodes are designed to be versatile, robust, reliable, accurate an

Strona 24

Ionode IJ Instruction Manual 30 4.7. Specifications Table 5: IJ44 pH electrode specifications Parameter Operating Range pH Range 0-14 pH (C glass)

Strona 25 - 4.5.3. Sleeve Electrolytes

Ionode IJ Instruction Manual 31 5. ORP 5.1. Theory ORP stands for oxidation reduction potential. Ideal ORP obeys the Nernst equation: 

Strona 26

Ionode IJ Instruction Manual 32 way pH does not characterize the buffering capacity. 5.1.1. Output ORP values are determined by measuring

Strona 27 - 4.6. Troubleshooting

Ionode IJ Instruction Manual 33 Table 6: Potential Relationships (mV) of Several Reference Electrodes at 25ºC SHE SCE Saturated KCl Ag/AgCl/ 1M

Strona 28

Ionode IJ Instruction Manual 34 5.2. Measurement Unlike a pH electrode, an ORP electrode does not require asymmetry potential and slope correct

Strona 29 - (>0.5pH)

Ionode IJ Instruction Manual 35 different potentials and irreversible redox couples. 4) Use only a high impedance mV meter. 5) Use fresh stan

Strona 30 - 4.7. Specifications

Ionode IJ Instruction Manual 36 5.4. Calibration Standards Light's solution and ZoBell's solution are NBS standards which are stable

Strona 31 - 5.1. Theory

Ionode IJ Instruction Manual 37 Table 8: ZoBell’sSolution vs Ag/AgCl/Sat KCl Reference) Temperature (ºC) Potential (mV) 5 273 10 262 15 251 20

Strona 32 - 5.1.1. Output

Ionode IJ Instruction Manual 38 5.5. Troubleshooting A fault may develop in either the meter or electrode. If a number of electrodes a

Strona 33

Ionode IJ Instruction Manual 39 Symptom Possible Causes Remedy Not within Tolerance (±20mV) Contaminated Standards Replace standards Correlating to

Strona 34 - 5.2. Measurement

Ionode IJ Instruction Manual 4 3. Preparation PLEASE NOTE: IJ series electrodes are shipped without sleeve electrolyte and must be filled prio

Strona 35 - 5.3. Maintenance

Ionode IJ Instruction Manual 40 5.6. Specifications Table 11: IJ64 and IJ64_xx specifications Parameter Operating Range Range -2000mV to +2000mV Acc

Strona 36 - 5.4. Calibration Standards

Ionode IJ Instruction Manual 41 6. Metal 6.1. Application – IJAu The IJAu uses a gold rod as sensor and is designed for determining ORP

Strona 37

Ionode IJ Instruction Manual 42 6.3. Specifications Table 12: IJAu and IJAg specifications Parameter Operating Range Range -2000mV to +2000mV Accur

Strona 38 - 5.5. Troubleshooting

Ionode IJ Instruction Manual 43 7. Optional Accessories RE45 - 3M KCl reference electrolyte gel,45mL KS200 –Electrode storage solution, 200ml KC200

Strona 39

Ionode IJ Instruction Manual 44 Warranty Any electrode found to be faulty due to manufacture will be replaced. All IJ electrodes have a serial numbe

Strona 40 - 5.6. Specifications

Ionode IJ Instruction Manual 5 out the end of the electrode through the ground junction. Check this has occurred. Insufficient electrol

Strona 41 - 6.2. Application – IJAg

Ionode IJ Instruction Manual 6 4. pH 4.1. Theory pH is a scale of active hydrogen ion concentration in aqueous solution. In pure water,

Strona 42 - 6.3. Specifications

Ionode IJ Instruction Manual 7 4.2. Electrode Characteristics 4.2.1. Construction A pH electrode consists of two electrochemical half-cells usuall

Strona 43 - 7. Optional Accessories

Ionode IJ Instruction Manual 8 Part of the circuit is electronic (heavy lines) (see Figure 6). The remainder is ionic. Unlike other pH electrode mea

Strona 44 - Warranty

Ionode IJ Instruction Manual 9 Unequal concentrations of H+ on either side of the membrane cause the system to equalise the number of positiv